![Hioki RM3542-50 Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/hioki/rm3542-50/rm3542-50_instruction-manual_2137361014.webp)
Operating Precautions
8
Follow these precautions to ensure safe operation and to obtain the full benefits of the various
functions.
Before using the instrument for the first time, verify that it operates normally to ensure that no
damage occurred during storage or shipping. If you find any damage, contact your authorized
Hioki distributor or reseller.
Installing
• To prevent overheating, be sure to leave the specified clearances around the instrument.
• Install the instrument with the bottom facing down.
Operating Precautions
Preliminary Checks
Instrument Installation
Installing the instrument in inappropriate locations may cause a malfunction of instru-
ment or may give rise to an accident. Avoid the following locations:
•
Exposed to direct sunlight or high temperature
•
Exposed to corrosive or combustible gases
•
Exposed to a strong electromagnetic field or electrostatic charge
•
Near induction heating systems (such as high-frequency induction heating systems
and IH cooking equipment)
•
Susceptible to vibration
•
Exposed to water, oil, chemicals, or solvents
•
Exposed to high humidity or condensation
•
Exposed to high quantities of dust particles
Do not place the instrument on an unstable table or an inclined place. Dropping or knocking
down the instrument can cause injury or damage to the instrument.
The instrument can be used with the stand (p. 16) in the upright position.
It can also be rack-mounted (p. A8).
50 mm or more
10 mm or more
Rear Panel
50 mm or more
Unplugging the power cord kills power to the instrument. Be sure to provide enough
unobstructed space to unplug the power cord immediately in an emergency.
Содержание RM3542-50
Страница 1: ......
Страница 2: ......
Страница 6: ...Table of Contents iv...
Страница 26: ...1 3 Screen Organization 20...
Страница 32: ...2 3 Turning the Power On and Off 26...
Страница 48: ...3 8 Confirming Faulty Measurements 42...
Страница 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Страница 84: ...5 7 Initializing Reset 78...
Страница 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Страница 206: ...11 4 Disposing of the Instrument 200...
Страница 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Страница 230: ...Index Index 4...
Страница 231: ...HIOKI E E CORPORATION...
Страница 232: ......
Страница 233: ......
Страница 234: ......