114
List of Error Messages
9.4 List of Error Messages
The table below lists messages displayed on the screen in the event of the product or measurement
state error. If repair is necessary, please contact your authorized Hioki distributor or reseller.
RM2611 Electrode Resistance Meter
Display
Meaning
Solution
NO UNIT
The MUX unit has not been
inserted.
Properly insert the MUX unit. If the
error persists, the instrument is
malfunctioning. Please have it repaired.
ERR:090
ROM check sum error.
Program ROM checksum error
The instrument is malfunctioning. Please
have it repaired.
ERR:091
RAM error.
CPU RAM error
The instrument is malfunctioning. Please
have it repaired.
ERR:092
Memory access failed.
Main power off, restart
after 10s.
Memory access error
Turn off the main power and wait at least
10 s before turning it back on.
ERR:093
Memory read/write
error.
Memory read/write test error
The instrument is malfunctioning. Please
have it repaired.
ERR:095
Adjustment data error.
Adjustment data error
The instrument is malfunctioning. Please
have it repaired.
ERR:096
Backup data error.
Configuration backup error
The settings have been initialized.
Reconfigure measurement conditions
and other settings.
ERR:097
Power line detection
error. Select power
line cycle.
Power supply frequency
detection error
The instrument was unable to detect
the power supply frequency. The power
line may be carrying significant noise.
Either move the instrument away from
the source of noise or affix a ferrite core
or other noise-reducing device to the
power cord.
ERR:098
Blown FUSE or
measurement lead is
broken.
The fuse is blown.
Replace the fuse. The fuse is not
customer-replaceable, so please contact
your authorized Hioki distributor or
reseller.
ERR:100
MUX unit error.
MUX unit error
The instrument is malfunctioning. Please
have it repaired.
Содержание RM2610
Страница 2: ......
Страница 12: ...8 Operating Precautions...
Страница 22: ...18 USB License Key and Screens RM2612 Resistance Calculation Software...
Страница 54: ...50 Checking Measurement Results...
Страница 72: ...68 Setting Administrator Mode...
Страница 110: ...106 Under what circumstances do contact errors occur...
Страница 128: ...124 Dimensional Drawings 10 2 Dimensional Drawings RM2611 Electrode Resistance Meter Unit mm...
Страница 130: ...126 Dimensional Drawings RM9005 Connection Cable Unit mm...
Страница 134: ...130 Index...
Страница 135: ......
Страница 136: ......
Страница 137: ......
Страница 138: ......