109
Peak Hold Function
5.4 Peak Hold Function
Pressing the
PEAK HOLD
key places the instrument in the peak hold state. Only parameters
whose values exceed the past peak value are refreshed. This function is useful to thoroughly
capture phenomena characterized by instantaneously large values, for example rush current.
During peak hold operation, the
PEAK HOLD
key is lit up in red, and the
[PEAK HOLD]
icon on
the screen’s operating status indicator appears.
See “Common screen display” (p. 23).
Peak hold
Refresh
Refresh
Displayed value
Internally-retained
measured value
Peak hold starts.
(Lit up in red)
Peak hold is released.
Time
Display value
When the past peak value is exceeded, the displayed value for that parameter is refreshed.
Measurement continues internally.
Canceling the peak hold state
Press the
PEAK HOLD
key again during the peak hold operation to cancel the peak hold state.
Various Functions
HIOKI PW8001A961-00
Содержание PW8001-01
Страница 2: ...HIOKI PW8001A961 00...
Страница 6: ...Contents iv HIOKI PW8001A961 00...
Страница 56: ...50 Checking Connections HIOKI PW8001A961 00...
Страница 120: ...114 Power Calculation Method HIOKI PW8001A961 00...
Страница 162: ...156 Controlling Integration with External Signals HIOKI PW8001A961 00...
Страница 236: ...230 U7005 15MS s Input Unit HIOKI PW8001A961 00...
Страница 258: ...252 Index HIOKI PW8001A961 00...
Страница 259: ...HIOKI PW8001A961 00...
Страница 260: ...HIOKI PW8001A961 00...
Страница 261: ...HIOKI PW8001A961 00...
Страница 262: ...www hioki com HIOKI PW8001A961 00...