PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
49/107
Execute and Clear Motor Analysis Function
’s Phase Zero Adjust
Syntax
Command
:EXTernalinA:PHASe:ZEROadjust <SET/CLEAR>
SET
: Phase zero adjust execution
CLEAR : Phase zero adjust value clear
Description
Command
Executes the phase zero adjust of the motor analysis function, or clear the
phase zero adjust value.
Same operation as the SHIFT+0ADJ from the main instrument, or the
SHIFT+DATA RESET key.
Example
Command
:EXTA:PHAS:ZERO SET
Execute the phase zero adjust.
Set and Query Motor Analysis Function Channel A Range
Syntax
Command
:EXTernalinA:RANGe <1/5/10>
Query
:EXTernalinA:RANGe?
Response
<1/5/10> Range value
Description
Command
Sets the voltage range of Channel A.
Query
Returns the range setting of Channel A as a numerical value.
Example
Command
:EXTA:RANG 10
Set Channel A to 10v range.
Query
:EXTA:RANG?
Response
:EXTERNALINA:RANGE 10 (WHEN HEADER IS ON)
10 (WHEN HEADER IS OFF)
Note
This command is only effective under the :EXTernalinA:FREQuency OFF.