PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
39/107
Set and Query Saved Items of Voltage Data
Syntax
Command
:DATAout:ITEM:U <CH 1>,< CH 2>,< CH 3>,< CH 4>,
<SUM12>,<SUM34>,<SUM123>
Query
:DATAout:ITEM:U?
Response
<CH 1>,< CH 2>,< CH 3>,< CH 4>,<SUM12>,<SUM34>,<SUM123>
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
CH 1
PK-
PK+
THD/RF
FND
DC
AC
MN
RMS
CH 2
PK-
PK+
THD/RF
FND
DC
AC
MN
RMS
CH 3
PK-
PK+
THD/RF
FND
DC
AC
MN
RMS
CH 4
PK-
PK+
THD/RF
FND
DC
AC
MN
RMS
SUM12
-
-
-
-
-
-
MN
RMS
SUM34
-
-
-
-
-
-
MN
RMS
SUM123
-
-
-
-
UNB
-
MN
RMS
Description
Command
Sets the saved voltage data items to the numerical value 0
– 255.
Query
Returns the saved voltage data items as numerical values.
Example
Command
:DATA:ITEM:U 1,1,1,1,0,0,0
Set the voltage RMSdata of Channels 1
– 4 to ON.
Query
:DATA:ITEM:U?
Response
:DATAOUT:ITEM:U 1,1,1,1,0,0,0 (WHEN HEADER IS ON)
1,1,1,1,0,0,0 (WHEN HEADER IS OFF)
Note
Data in the THD/RF items changes according to the integration mode setting.
Some items are unselectable due to the wiring and integration mode.
When an unselectable item is specified, an execution error will occur and other items will also
not be set.
Use this command after :DATAout:ITEM:NORMal and :DATAout:ITEM:SUMCommand.
Содержание PW3390