PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
34/107
Set and Query Output Order of Harmonic Data Saving
Syntax
Command
:DATAout:ITEM:HARMonic:ORDer <Lower limit order>,
<Upper limit order>,<ODD/EVEN/ALL>
Query
:DATAout:ITEM:HARMonic:ORDer?
Response
Lower limit order (NR1)
:0 - 100
Upper limit order (NR1)
:0 - 100
ODD
: Odd-number order only
EVEN
: Even-number order only
ALL
: All orders
Description
Command
Sets the upper limit, lower limit orders, even-number,
odd-number and all orders for the harmonic saved items.
Use in combination with
":DATAout:ITEM:HARMonic:LIST"Command.
Query
Returns the setting for the order of the harmonic saved items with numerical
and character string.
Example
Command
:DATA:ITEM:HARM:ORD 1,15,ODD
Set the odd-number order from 1
– 15 to the default output.
Query
:DATA:ITEM:HARM:ORD?
Response
:DATAOUT:HARMONIC:ORDER 1,15,ODD (WHEN HEADER IS ON)
1,15,ODD (WHEN HEADER IS OFF)