PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
31/107
Set and Query Efficiency, Saved Items of Loss Calculation Value
Syntax
Command
:DATAout:ITEM:EFFiciency <0 to 255>
Query
:DATAout:ITEM:EFFiciency?
Response
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
-
LOSS3 LOSS2 LOSS1
-
EFF3
EFF2
EFF1
Description
Command
Sets the efficiency, saved items of loss calculation values between
0 and 255.
Query
Returns the setting of efficiency, saved items of loss calculation values with
values of 0
– 255.
Example
Command
:DATA:ITEM:EFF 17
Set
“1” of efficiency calculation value, and “1” of loss calculation value to ON.
Query
:DATA:ITEM:EFF?
Response
:DATAOUT:ITEM:EFF 17 (WHEN HEADER IS ON)
17 (WHEN HEADER IS OFF)
Set and Query Saved Items of Motor Input
Syntax
Command
:DATAout:ITEM:EXTernalin <0 to 255>
Query
:DATAout:ITEM:EXTernalin?
Response
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
-
-
-
-
SLIP
PM
EXTB
EXTA
Description
Command
Sets motor saved items between 0 and 255.
Query
Returns setting of motor saved items with values of 0
– 255.
Example
Command
:DATA:ITEM:EXT 7
Set the EXTA, EXTB, PM of motor analysis function calculation value to ON.
Query
:DATA:ITEM:EXT?
Response
:DATAOUT:ITEM:EXTERNALIN 7 (WHEN HEADER IS ON)
7 (WHEN HEADER IS OFF)
Содержание PW3390