4.2 Changing Measurement Settings
63
3
Chapter 4 Changing Set
tings
4
Voltage range
The voltage range is fixed to 600 V.
VT ratio (PT ratio)
Set when using a VT (PT) to perform measurement.
Measurement 2 Setting Screen
Selection
Manual
0.01 to 9999.99
Select
1/60/100/200/300/600/700/1000/2000/2500/5000
• When taking measurements on the secondary side of a voltage
transformer (VT), you can set the VT ratio in order to convert the
readings to their primary-side equivalents and display the
results. For a VT with a primary-side voltage of 6.6 kV and a sec-
ondary-side voltage of 110 V, the VT ratio would be 60 (6,600 V /
110 V).
• Since the current range is fixed at 600 V, it would be multiplied by
the VT ratio of 60 to yield a voltage range of 36 kV.
Содержание PW3360-20
Страница 2: ......
Страница 66: ...3 8 Verifying Correct Wiring Wiring Check 58 ...
Страница 104: ...7 2 Settings That Can Be Added to Quick Set Settings 96 ...
Страница 120: ...8 8 Formatting the SD Memory Card or Internal Memory 112 ...
Страница 166: ...11 4 Outputting a Pulse Signal 158 ...
Страница 210: ...13 4 Disposing of the Instrument 202 ...
Страница 218: ...A8 ...
Страница 222: ...Index Index 4 ...
Страница 223: ......
Страница 224: ......