3.11 When PEAK OVER, o.r, or the Unit Indicator Flashes
116
3.11.3 When the Unit Indicator Flashes
Status
Solution
The unit indicator flashes during
averaging processing
The average value being displayed contains [
o.r
] data. If no [
o.r
] data is present during
averaging processing, the unit indicator will not flash. When [
o.r
] is displayed, internal
data for which the voltage or current exceeded 140% of the range, or the active power
exceeded 196% of the range, will be used as-is in calculating the average value.
The integrated value or time aver-
age value unit indicator flashes
Reset integrated values, change the range, and repeat integration. If no
PEAK OVER
condition occurs during integration, the indicator will not flash.