2.6 Connecting Measurement Cables to the Input Terminals
20
Connect the measurement cables to the measurement terminals on the rear of the instrument.
Wires with either crimp terminals or banana-plug cables can be connected.
How to connect cables
2.6 Connecting Measurement Cables to the
Input Terminals
Before connecting cables, be sure to read "Connecting Cables" (p. 8) and "Handling the
Cords and Cables" (p. 7).
•
The maximum input voltage is 250 VDC. Attempting to measure voltage in excess
of the maximum input could destroy the instrument and result in personal injury or
death.
•
The maximum rated voltage between input terminals and ground is 300 VAC, DC.
Attempting to measure voltages exceeding 300 V with respect to ground could
damage the instrument and result in personal injury.
To avoid experiencing an electric shock or causing a short-circuit at the input termi-
nals, only insulated crimp contacts should be used on wires being connected to
input pins. (The input terminal pin diameter is M6.)
CH1
(PR8112)
The PR8111 does not have terminals for CH2.
• When you input a signal such that the red (H) terminal has a higher potential than the black (L)
terminal, the pen moves rightward, starting from the user-specified zero point.
• When you input a signal such that the red (H) terminal has a lower potential than the black (L)
terminal, the pen moves leftward, starting from the user-specified zero point.
CH2
Содержание PR8111
Страница 1: ...PR8111 PR8112 Instruction Manual PEN RECORDER EN Feb 2020 Revised edition 3 PR8111A981 03 20 02H...
Страница 2: ......
Страница 4: ...Contents ii...
Страница 32: ...3 4 Example Printout 28...
Страница 36: ...Chapter 4 Specifications 32...
Страница 41: ...1...
Страница 42: ......
Страница 43: ......
Страница 44: ......