5.3 Changing the Measurement Period
80
This section describes how to set the method used to start and stop recording.
5.3 Changing the Measurement Period
Display a
pull-down menu
to select a setting
Select a value
to change
Select the setting/
change the value
Accept setting
Cancel
[Record]
[SYSTEM]
screen
Select a setting
Key operation during configuration
[Time]
Time Start
Setting Contents:(
∗
: Default setting)
Manual
Starts and stops recording when the
START/STOP
key is pressed.
Time
Starts/stops recording at the set time and date.
If the start time and date has already passed when the
START/STOP
key is pressed, an
[Exactly]
start will result, and the start time and date will automatically be changed to the present
time. If the stop time and date has also passed, the stop time and date will be repeated and auto-
matically changed to the longest recording setting interval.
Exactly
∗
Starts recording at the next
[TIME PLOT Interval]
after the
START/STOP
key is pressed.
If the
START/STOP
key is pressed at the time “10:41:22” with the time plot interval set to 10 min,
the instrument will enter in the standby state. The recording will start at “10:50:00.”
If the set recording interval is of 30 s or less, recording will start from the next zero second.
Stop operation is the same as with the
[Manual]
setting.
HIOKI PQ3198A961-03
Содержание PQ3198
Страница 2: ...HIOKI PQ3198A961 03...
Страница 6: ...Contents iv HIOKI PQ3198A961 03...
Страница 20: ...Usage Notes 14 HIOKI PQ3198A961 03...
Страница 44: ...2 3 Display Items and Screen Types 38 HIOKI PQ3198A961 03...
Страница 58: ...3 8 Turning the Power On and Off Setting the Default Language 52 HIOKI PQ3198A961 03...
Страница 102: ...5 8 Factory Settings 96 HIOKI PQ3198A961 03...
Страница 116: ...6 4 Displaying Harmonics 110 Holding the display Hold Measured values and waveforms will be held HIOKI PQ3198A961 03...
Страница 198: ...12 4 Downloading Recorded Data to Computer 192 HIOKI PQ3198A961 03...
Страница 264: ...14 4 Disposing of the Instrument 258 HIOKI PQ3198A961 03...
Страница 296: ...A32 HIOKI PQ3198A961 03...
Страница 299: ...Index Index 3 Index Wiring 73 Z Zero adjustment 53 Zero phase A28 HIOKI PQ3198A961 03...
Страница 300: ...Index Index 4 HIOKI PQ3198A961 03...
Страница 301: ...HIOKI PQ3198A961 03...
Страница 302: ...HIOKI PQ3198A961 03...
Страница 303: ...HIOKI PQ3198A961 03...
Страница 304: ...www hioki com HIOKI PQ3198A961 03...