13.3 Screen Specifications
221
10
13
Cha
p
ter 13 Spe
cificatio
ns
(7) Easy settings
Abnormal voltage
detection
Basic power supply
quality measurement
Inrush current
measurement
Measured value
recording
EN50160
Connection
Set in advance
Current sensor
Set in advance
CT, PT ratios
Set in advance
Measurement frequen-
cy
Automatic detection of 50 Hz/60 Hz/400 Hz; if unable to detect, user (manual) setting
Nominal input voltage
Automatic detection; if unable to detect, user (manual) setting
Flicker
Pst, Plt
Pst, Plt
Pst, Plt
Pst, Plt
Pst, Plt
Measurement RMS
voltage selection
Default
Default
Default
Default
Default
Measurement
harmonics selection
RMS value
RMS value
RMS value
RMS value
Content percentage
Total harmonic
distortion factor selection
THD_F
THD_F
THD_F
THD_F
THD_F
Power factor selection PF
PF
PF
PF
PF
Repeat setting and
iterations
OFF
(max. 35 days)
OFF
(max. 35 days)
OFF
(max. 35 days)
OFF
(max. 35 days)
OFF
(max. 35 days)
Recorded items setting Power and Harmon-
ic
All data
Power and Harmon-
ic
All data
All data
TIME PLOT interval
1 minute
10 minutes
1 minute
10 minutes
10 minutes
Current range
Automatic detection Automatic detection Max. range
Automatic detection Automatic detection
Event hysteresis
1%
1%
1%
1%
2%
Transient overvoltage
70% of nominal
voltage
70% of nominal
voltage
OFF
OFF
100% of nominal
voltage
Voltage swell
110% of nominal
voltage
110% of nominal
voltage
OFF
OFF
110% of nominal
voltage
Voltage dip
90% of nominal
voltage
90% of nominal
voltage
OFF
OFF
90% of nominal
voltage
Interruption
10% of nominal
voltage
10% of nominal
voltage
OFF
OFF
1% of nominal
voltage
Frequency 200 ms
±5 Hz of nominal
frequency
±0.5 Hz of nominal
frequency
OFF
OFF
±0.5 Hz of nominal
frequency
Frequency cycle
OFF
OFF
OFF
OFF
OFF
Voltage waveform
peak (±)
150% of reference
value
150% of reference
value
OFF
OFF
170% of nominal
voltage
Voltage DC fluctuation
(±) (when DC is selected)
±10% based on DC
measured value
±10% based on DC
measured value
OFF
OFF
OFF
Current waveform
peak (±)
OFF
200% of reference
value
300% of reference
value
OFF
OFF
Current DC fluctuation (±)
(when DC is selected)
±10% based on DC
measured value
±10% based on DC
measured value
OFF
OFF
OFF
RMS voltage
10% of reference
value
SENSE width: ±10 V
10% of reference
value
SENSE width: ±10 V
OFF
OFF
OFF
RMS current
OFF
SENSE width: OFF
50% of reference
value
SENSE width: OFF
OFF
SENSE width: OFF
OFF
SENSE width: OFF
OFF
SENSE width: OFF
Inrush current
(Irms 1/2)
OFF
OFF
200% of reference
value
OFF
OFF
Active power
OFF
OFF
OFF
OFF
OFF
Apparent power
OFF
OFF
OFF
OFF
OFF
Reactive power
OFF
OFF
OFF
OFF
OFF
Power factor/displace-
ment power factor
OFF
OFF
OFF
OFF
OFF
Voltage unbalance
factor (zero-phase,
negative-phase)
OFF, 3%
OFF, 3%
OFF, OFF
OFF, OFF
OFF, 2%
Setting
Pattern
HIOKI PQ3198A961-03
Содержание PQ3198
Страница 2: ...HIOKI PQ3198A961 03...
Страница 6: ...Contents iv HIOKI PQ3198A961 03...
Страница 20: ...Usage Notes 14 HIOKI PQ3198A961 03...
Страница 44: ...2 3 Display Items and Screen Types 38 HIOKI PQ3198A961 03...
Страница 58: ...3 8 Turning the Power On and Off Setting the Default Language 52 HIOKI PQ3198A961 03...
Страница 102: ...5 8 Factory Settings 96 HIOKI PQ3198A961 03...
Страница 116: ...6 4 Displaying Harmonics 110 Holding the display Hold Measured values and waveforms will be held HIOKI PQ3198A961 03...
Страница 198: ...12 4 Downloading Recorded Data to Computer 192 HIOKI PQ3198A961 03...
Страница 264: ...14 4 Disposing of the Instrument 258 HIOKI PQ3198A961 03...
Страница 296: ...A32 HIOKI PQ3198A961 03...
Страница 299: ...Index Index 3 Index Wiring 73 Z Zero adjustment 53 Zero phase A28 HIOKI PQ3198A961 03...
Страница 300: ...Index Index 4 HIOKI PQ3198A961 03...
Страница 301: ...HIOKI PQ3198A961 03...
Страница 302: ...HIOKI PQ3198A961 03...
Страница 303: ...HIOKI PQ3198A961 03...
Страница 304: ...www hioki com HIOKI PQ3198A961 03...