35
How to Use Voltage Sensors
Improperly affixed sensor
A voltage sensor affixed improperly will be susceptible to effects of nearby
cables, resulting in inaccurate measurement.
The insulated cable is pinched
by the tips of the
clip.
The sensor is affixed to the
insulated cable at an angle.
Insulated cable
Insulated cable
The insulated cable is inserted
too close to the pivot point.
The sensor is affixed to two
insulated cables together, each of
which carries a voltage different
from each other.
Insulated cable
Insulated cable
IMPORTANT
If an insulated cables under measurement are dirt or wet, the instrument
may display values lower than the actual voltages. In this case, wipe them
clean with a dry cloth before measurement.