![Hioki MR8904 Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/hioki/mr8904/mr8904_instruction-manual_2137100016.webp)
3.2 Connecting the Unit to the Measurement Target
12
Before connecting the unit, be sure to read "When Connecting the MR8904 and the Object under Mea-
surement" (p.5). Then connect the MR8904 to the measurement target with a CAN cable.
Required items: Model 9713-01 CAN Cable, Phillips head screwdriver (No.1)
3.2
Connecting the Unit to the Measurement
Target
Example: Connecting the unit with the 9713-01
Wire
color
Pin no.
Receive
signal name
-
1
NC
Green
2
CAN_L
Black
3
CAN_GND
- 4
NC
-
5
NC
- 6
NC
Red
7
CAN_H
- 8
NC
White
9
CAN_V+
- Shield
GND
1
2
3
4
5
6
7
8
9
Connector
D-Sub 17LE23090-27
(Manufactured by DDK Ltd.)
Top
Screw
1
2
1
Connect the CAN cable to one of
the MR8904 CAN Unit’s ports.
Take care to orient the connector
properly. Tighten the fixing screw
with a Phillips head screwdriver to
secure the connector to the unit.
2
Attach to the measurement object.