4.3 Using High-speed Function (Instantaneous Recording)
71
Chapter 4
Set
ting t
h
e Meas
urement
Condit
ions
(
W
hen you want t
o
c
u
st
omize t
h
e
sett
ings)
4
Two setting procedures are available: setting the time
per division of the horizontal axis and setting the data
acquisition interval.
• When setting the time per division of the horizontal
axis
The data of 100 samples is included in 1 division.
(When the magnification is x1)
Sampling rate = timebase (seconds/div)/100 (data
samples/division)
• When setting the data acquisition interval (p. 72)
(It is possible to switch to the same setting proce-
dure as for the Real-time Function.)
Select the recording interval that matches the mea-
surement target. If a short recording interval is selected, the recording time for
which data can be stored in the internal buffer memory will be short.
Set the time per division
Setting the Horizontal Axis (Time Base Range)
1
division
1
100
Data acquisition interval
Time per division
Setting options (
∗
. default setting)
• Decreasing the timebase increases the resolu-
tion of data available for analysis.
See:
"Selecting a Timebase (Horizontal Axis)" (p.
A19)
• When the product of the timebase multiplied by
the display magnification is greater than 50 ms,
waveforms can be automatically scrolled on the
display while measuring (the Roll Mode display
function).
100
∗
, 200, 500
μ
s
[/div]
1, 2, 5, 10, 20, 50, 100 ms
[/div]
The setting can also be
changed on the Waveform
screen. ([Wave+Set] Dis-
play)
1
2
Select from the setting options of
[Timebase]
.
(Open/confirm the setting information.)
Select the
[Setting]
screen.
Содержание MR8880-20
Страница 2: ......
Страница 12: ...Procedure 2 ...
Страница 40: ...1 3 Screen Configuration and Display 30 ...
Страница 76: ...3 4 Measurement Using Setting Wizard PRESETS key 66 ...
Страница 174: ...6 4 Data Management 164 ...
Страница 186: ...7 7 Performing a Printer Check 176 ...
Страница 196: ...8 4 Numerical Value Calculation Expressions 186 ...
Страница 216: ...10 4 Making System Settings 206 ...
Страница 222: ...11 3 Communication Using Commands 212 4Click Next to start install ing Installing Click ...
Страница 226: ...11 3 Communication Using Commands 216 ...
Страница 254: ...13 6 Setting Wizard Function PRESETS 244 ...
Страница 292: ...Appendix 10 Installing Waveform Viewer A30 ...
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......