![Hioki MR8880-20 Скачать руководство пользователя страница 229](http://html1.mh-extra.com/html/hioki/mr8880-20/mr8880-20_instruction-manual_2137332229.webp)
12.2 External Signal Input (External Trigger Input)
219
Chapter 12 Ext
ernal Cont
rol
12
Triggering can be controlled by applying a signal from an external trigger source (p.
127). Furthermore, an event mark can be added by an input signal (p. 142).
This allows synchronous operation of multiple Memory HiCorders by parallel trigger-
ing (p. 226).
12.2 External Signal Input (External Trigger Input)
External signal
input (EXT.TRIG)
Trigger event occurs.
External
Trigger Input Signals
Voltage range
HIGH level: 2.5 to 5.0 V, LOW level: 0 to 0.8 V
Pulse width
• Filter Off:
HIGH level: 1 ms or greater, LOW level: 2
μ
s or greater
• Filter On:
HIGH level: 2.5 ms or greater, LOW level: 2.5 ms or greater
Maximum input voltage
0 to 10 V DC
100 pF
3.3 k
Ω
5 V
5 V
GND
EXT.TRIG
HIGH
2.5 to 5.0 V
LOW
0 to 0.8 V
[
↓
]
[
↓
]
1 ms or greater (Filter Off)
2.5 ms or greater (Filter On)
2
μ
s or greater (Filter Off)
2.5 ms or greater (Filter On)
1 k
Ω
When using an external trigger to synchronize multiple instruments, make
sure you set
[Filter]
of the external trigger input to
[Off]
.
Содержание MR8880-20
Страница 2: ......
Страница 12: ...Procedure 2 ...
Страница 40: ...1 3 Screen Configuration and Display 30 ...
Страница 76: ...3 4 Measurement Using Setting Wizard PRESETS key 66 ...
Страница 174: ...6 4 Data Management 164 ...
Страница 186: ...7 7 Performing a Printer Check 176 ...
Страница 196: ...8 4 Numerical Value Calculation Expressions 186 ...
Страница 216: ...10 4 Making System Settings 206 ...
Страница 222: ...11 3 Communication Using Commands 212 4Click Next to start install ing Installing Click ...
Страница 226: ...11 3 Communication Using Commands 216 ...
Страница 254: ...13 6 Setting Wizard Function PRESETS 244 ...
Страница 292: ...Appendix 10 Installing Waveform Viewer A30 ...
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......