12.1 Connecting to the External Control Terminals
217
Chapter 12 Ext
ernal Cont
rol
12
The external control terminals on the Memory HiCorder support trigger signal input
and output.
External Control
Chapter 12
1
START/IN1
Input signals for controlling the operation of the
instrument.
(Start, stop, abort, print)
Input
2
STOP/IN2
3
GND
(common with instrument ground)
4
ALM/OUT
Output signals for instrument status.
(Printing, saving, instrument alarm)
Output
5
TRIG.OUT
Output signal when trigger applied.
Output
6
EXT.TRIG
(External trigger input)
Input external signals.
Input
External control
terminals
(Right side)
12.1
Connecting to the External Control
Terminals
To avoid electrical hazards and damage to the instrument, do not
apply voltage exceeding the rated maximum to the external con-
trol terminals.
I/O terminals
Maximum input voltage
Input
EXT.TRIG
0 to 10 V DC
START/IN1
STOP/IN2
Output
ALM/OUT
0 to 50 V DC
TRIG.OUT
Содержание MR8880-20
Страница 2: ......
Страница 12: ...Procedure 2 ...
Страница 40: ...1 3 Screen Configuration and Display 30 ...
Страница 76: ...3 4 Measurement Using Setting Wizard PRESETS key 66 ...
Страница 174: ...6 4 Data Management 164 ...
Страница 186: ...7 7 Performing a Printer Check 176 ...
Страница 196: ...8 4 Numerical Value Calculation Expressions 186 ...
Страница 216: ...10 4 Making System Settings 206 ...
Страница 222: ...11 3 Communication Using Commands 212 4Click Next to start install ing Installing Click ...
Страница 226: ...11 3 Communication Using Commands 216 ...
Страница 254: ...13 6 Setting Wizard Function PRESETS 244 ...
Страница 292: ...Appendix 10 Installing Waveform Viewer A30 ...
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......