74
Start and Stop Measurement
4.3 Start and Stop Measurement
Press the
DISP
key to go to the Waveform screen.
1
Start measurement.
Press the
START
key to start measurement.
2
Set the pen to Up/Down.
This setting can be chosen before or during measurement. When the pen is set to Down, the waveform
is being drawn. When the pen is set to Up, the waveform is not drawn. This setting applies to all
channels.
Move the cursor to
[Pen]
to change the setting.
Up
Waveform is not drawn.
Down
Waveform is drawn.
Up/Down operation is also possible
using the
TRIG.SET
key or the
Manual
Trigger
key.
Pen in Down position
Pen is shown in color and waveform is drawn.
Pen in Up position
Pen is shown in white and waveform is not drawn.
3
Stop measurement
Press the
STOP
key to stop measurement.
4
Redraw waveforms
When measurement is stopped, waveform data for up to 16,000,000 samples before the current point can
be played back (redrawn) by the pen, similar to using a video recorder, starting from any specified point.
This setting applies to all channels. The playback speed can be specified. (It can also be changed during playback.)
Waveform is only played back on the screen and will not affect the printout data. To print the same
waveforms as shown on the screen, use the hard copy function (p. 118).
Move the cursor to
[Player]
.
Clear
Clear only waveform display.
(Waveform data will not be cleared.)
Redraw
Redraw measurement data. Measurement
settings can also be changed for redrawing.
First
Move pen to start of waveform.
Last
Move pen to end of waveform.
Play/Stop
Start/stop waveform playback.
Move the cursor to
[Speed]
.
With the
[Normal]
setting, playback
occurs at the same speed as recording.
Very Fast/ Fast /Normal
(Default setting)/ Slow/Very Slow
Measurement start/stop and pen up/down can be controlled via the external control terminal. (p. 338)
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......