69
Start and Stop Measurement
Measurement and Internal Operations
Measurement methods are normal measurement (start recording when measurement starts) and trigger
measurement (start recording when trigger criteria are satisfied). In this manual, “
Measurement start
”
means when you press the
START
key and “
Recording start
” means when recording begins on the
waveform screen.
•
To record on either single or repeating trigger events
Select the Trigger mode.
•
To capture data measured prior to trigger events
Enable pre-triggering.
Normal
Measurement
Without triggering
START
key
End of Measurement
Recording
Starts
Recording
Stops
[Now Storing]
[Storing Finished]
[ ]: Status Bar Display
Recording
Start Measurement
Terminates when measurement up to the recording length has made.
Trigger
Measurement
Trigger modes:
[Single]
Pre-triggering not
enabled
Single triggering
End of Measurement
[Trigger Wait]
Recording
Starts
T
Recording
Stops
[Now Storing]
Start Measurement
START
key
[Storing Finished]
Recording
Recording starts when a trigger event occurs and continues for the specified recording
length.
Repeated triggering
Trigger modes:
[Repeat]
Pre-triggering not
enabled
Stop Measurement
[Trigger Wait]
Recording
Starts
T
Recording
Stops
[Now Storing]
Start Measurement
Recording
[Trigger Wait]
T
Recording
Starts
START
key
STOP
key
Recording starts when a trigger event occurs, continues for the specified recording length,
and returns to the Trigger Wait state.
Repeated triggering and recording of phenomena before each event
Trigger modes:
[Repeat]
Pre-triggering
enabled
[Pre-Trigger Wait]
End of Measurement
[Trigger Wait]
Start Measurement
T
Recording
Stops
[Now Storing]
Recording
Starts
The specified pre-trigger wait period is recorded before each trigger event
Recording
[Storing Finished]
T
Waveform during
specified pre-trigger
period
Post-trigger waveform
Operation is
repeated from
pre-triggering
stage.
START
key
After starting measurement and internally acquiring data for the
specified pre-trigger period, the
Trigger Wait
state is activated.
The data before a trigger event (for the pre-trigger period) is recorded.
When the trigger mode is set to
[Repeat]
or
[Auto]
(Memory function only), the
number of trigger events is shown in the top part of the screen (Storage Counter).
Refer to “Explanation of Screen Contents” (p. 20).
3
Measurement
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......