360
Module Specifications
18.6
Module Specifications
18.6.1
Model 8966 Analog Unit
The accuracy is provided for the condition where the module is installed on the Memory HiCorder at 23°C±5°C
(73°F±9°F), 20% RH to 80% RH, and zero adjustment is executed 30 minutes after power on.
Product warranty
period
3 years
Guaranteed accuracy
period
1 year
Number of input
channels
2 channels
Measurement range
5, 10, 20, 50, 100, 200, 500 mV, 1, 2, 5, 10, 20 V/div
Measurement
accuracy
±0.5% f.s. (Filter 5 Hz On)
Temperature
characteristic
±0.06% f.s./°C
Frequency
characteristic
DC Coupling: DC to 5 MHz −3 dB
AC Coupling: 7 Hz to 5 MHz −3 dB (low cut-off frequency: 7 Hz±50%)
Noise
1.5 mV p-p (typ.), 2 mV p-p (max.) (highest sensitivity range, with input shorted)
Common mode
rejection ratio
80 dB or more (at 50 Hz/60 Hz and with signal source resistance 100
Ω
or less)
Low-pass filter
OFF, 5±50%, 50±50%, 500±50%, 5 k±50%, 50 k±50%, 500 k±50% (Hz) −3 dB
Input type
Unbalanced input (floating)
Input coupling
AC/ DC/ GND
Input resistance
1 M
Ω
±1%
Input capacitance
30 pF±10 pF (at 100 kHz)
A/D resolution
12 bits
Maximum sampling
rate
20 MS/s
Input terminals
Insulated BNC terminal
Maximum input
voltage
400 V DC
Maximum rated
voltage to earth
300 V AC, DC (between individual input channels and enclosure as well as
between individual input channels)
Measurement category II (anticipated transient overvoltage: 2500 V)
Operating
temperature and
humidity
As per the Memory HiCorder in which Model 8966 is installed
Operating
environment
As per the Memory HiCorder in which Model 8966 is installed
Storage temperature
and humidity
-10°C to 50°C (14°F to 122°F), 80% RH or less (no condensation)
Dimensions
Approx. 106W × 19.8H × 196.5D mm (4.17″W × 0.78″H × 7.74″D)
Mass
Approx. 250 g (8.8 oz.)
Effect of radiated
radio-frequency
electromagnetic field
±15% f.s. (max.) at 3 V/m
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......