Display Settings
Difference Between Dead Times During Normal and Memory Division Recording
When both printer recording (Auto Print) and Auto Save are set for continuous triggering
[Repeat]
Dead Times
Times during which sampling is inhibited due to internal processing, printing or saving
Recording
Length
Anomalous phenomena occurring during dead times are not detected.
When the Trace Waveform Display is disabled (
[Off]
) during Memory Division recording
1
2
3
4
5
6
Recording
Length
Dead Times
The waveform data of each recording length is recorded in one block.
When recording with Memory Division, dead time is shorter than with normal recording.
• The dead time (the period where no sampling occurs between blocks) of memory division is as
follows.
5 μs/div to 20 μs/div: 1 to 8 samples
Timebase slower than 50 μs/div: 1 sample
Note: Dead time is longer when you use calculation functions or when the timebase is set between 5
and 20 μs/div and tracking waveform display is
[On]
.
•
When measuring with an Model 8970 Freq Unit, dead time is approximately 230 ms. When measuring
in integral value ([Count]) mode, there are cases when the last data in a previous block is carried over
to the first part of the next block.
• When tracking waveform display is
[Off]
, the roll mode function cannot be used, even when roll mode
is enabled (not
[Off]
).
• When triggering occurs very often, pressing the
STOP
key may not stop measurement until enough
data has been acquired to fill the blocks specified for use.
Auto Save and Auto Print for Memory Division Recording
Measurement conditions Auto-save and Auto-print
With Numerical calculation
ON
Auto-save and Auto-print are performed every time one block is measured. When
Tracking wave display is ON, the Waveform is also displayed.
When timebase is 5 μs/div
to 20 μs/div and Tracking
wave display is ON.
Auto-save, auto-print and Waveform Display are performed every time one block is
measured.
When timebase is 5 μs/div
to 20 μs/div and Tracking
wave display is OFF.
Auto-save and Auto-print are performed after all blocks are measured.
Other than above
Auto-save and Auto-print are performed simultaneously with measurement. When
Tracking wave display is ON, the Waveform is also displayed.
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......