212
Numerical Value Calculation Settings
10.2 Numerical Value Calculation Settings
Procedure
Press the
STATUS
key to open the Status screen, and then select the
[Num Calc]
sheet.
1
Enable the Numerical Calculation function.
Move the cursor to
[Numerical Calc]
.
Select
[On]
.
1
2
2
Specify the waveform calculation range.
Move the cursor to
[Calc Area]
.
Whole Wave
Applies calculations to the whole waveform.
(Default setting)
A-B Wave
Applies calculations to the data between AB
cursors.
After Trigger
Applies calculations to the waveform after trigger.
Refer to “7.2 Specify a Waveform Range (AB Cursor)” (p. 125).
3
Choose calculation settings.
Calculation No.
(5)
Phenomenon
(1)
Calculation
Type
(2)
Channel for
Calculation
(4)
Statistical
calculation
(3)
Parameters*
*:
Setting choices depend
on the calculation type.
(1)
Select the calculation type.
Move the cursor to the number of the calculation type for which
you want to choose settings. Select the calculation type.
(When
[List]
is selected, a list of calculation types appears. To
clear the display, press the
TRIG.SET
key.)
Select [Judge] if you require
judgment of calculation results.
Off
No calculation.
(Default setting)
Area
Area enclosed by zero position and signal
waveform
Average
Average valu
x
e of waveform data
X-Y Area
Area of X-Y composite waveform
Rms
RMS value of waveform data
Time to Lev*
Time from trigger to specified level
Peak-Peak
Peak-to-peak value of waveform data
Lev-time
Measurement value at a specified time point after
triggering
Maximum
Maximum value of waveform data
Max-Time
Time from trigger to maximum value
Pulse Width*
Pulse width of waveform data
Minimum
Minimum value of waveform data
Duty Ratio*
Duty of waveform data
Min-Time
Time from trigger to minimum value
Pulses*
Pulse count of waveform data
Period*
Period of signal waveform
Calculation
Four arithmetic operations on numerical
calculation results
Frequency*
Frequency of signal waveform
Time Diff*
Time difference between phenomenon A and B.
RiseTime
Rise time of waveform data
Phase Diff*
Time difference between phenomenon A and B
displayed as a phase differnce.
FallTime
Fall time of waveform data
High Level
High level value for waveform data
Std Dev
Standard deviation of waveform data
Low Level
Low level value for waveform data
*: This setting can also be made for logic channels
When selecting
[A-B Wave]
, specify
the calculation range using the AB
cursors on the Waveform screen.
If no measurement data has been
acquired by the instrument, first
measure once so that the range
can be specified for calculations
to be applied to subsequent
measurements.
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......