209
10 Numerical Calculation Functions
10
Numerical Calculation
Functions
Numerical calculations can only be used with the Memory function.
Results calculated from the acquired waveform are displayed as numerical values on the Waveform
screen. These calculation results can be judged for evaluation. Numerical calculation settings are made
on the Status screen -
[Num Calc]
sheet.
To Open the
[Num Calc]
Sheet
Press this key repeatedly to
display the various sheets.
[Status]
[Num Calc]
Operations Available From the [Num Calc] Sheet
Numerical Calculations
•Average
•RMS (Root-Mean-Square) value
•Peak-to-Peak Value (P-P Value)
•Maximum Value
•Time to Maximum Value
•Minimum Value
•Time to Minimum Value
•Period
•Frequency
•Rise Time
•Fall Time
•Standard Deviation
•Area
•X-Y Area
•Time to Level
•Level at Time
•Pulse Width
•Duty Ratio (%)
•Pulse Count
•Four Arithmetic Operations
•Time difference calculation
•Phase differnce calculation
•High level calculation
•Low level calculation
(24 types in total)
•
Specified calculation between AB
cursors
Waveform calculations can be
limited to data within the range
specified by the AB cursors.
Calculation operator details: “10.6
Numerical Calculation Type and
Numerical Calculation Judgment (p. 216)
The result of a numerical calculation can be
compared to a preset reference range, for GO/NG
(Pass/Fail) evaluation.
There are 24 types of numerical calculations in total,
16 of which can be performed simultaneously.
When the Scaling function is used, numerical
calculation is performed on scaled values.
Save/Print Numerical Calculation Results
•Auto-save numerical calculation results
•
Save specified existing numerical calculation result
Refer to “10.4 Save Numerical Calculation Results” (p. 219).
•Auto-print
•Arbitrary printing
Refer to “10.5 Print Numerical Calculation Results” (p. 221).
10
Numerical Calculation Functions
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......