Trigger by Timer or Time Interval (Timer Trigger)
9.5 Trigger by Timer or Time Interval (Timer Trigger)
This trigger allows you to record at fixed times.
•
Triggering occurs at the specified interval from the specified Start time until the Stop time.
•
Before setting up a Timer Trigger, verify that the clock is set to the correct time. If not, set the clock on
the System screen -
[Init]
Procedure
Press the
DISP
key to open the Waveform screen, and then press the
TRIG.SET
key to open the trigger settings
window.
1
Enable or disable the timer trigger.
Move the cursor to the
[Timer Trig]
item.
Off
Timer triggering disabled.
On
Timer triggering enabled.
1
2
3
Start Time
Stop Time
T
T
T
Interval
Records for the specified recording length
START
key
2
(When
[On]
is selected)
Set Start and Stop times.
Move the cursor to
[M]
,
[D]
,
[h]
and
[m]
to set recording
Start and Stop times.
Set the date and time.
To set to the current date and time, select
[Present Time]
.
3
(To apply a trigger through the specified interval, from
Start to Stop)
Set the Interval.
Move to cursor to
[D]
,
[h]
,
[m]
and
[s]
of
[Interval]
.
Set the recording interval.
4
Press the
START
key.
Measurement starts and recording will start at the specified
Start time.
To stop recording early, press the
STOP
key.
When the specified interval
is shorter than the specified
recording length:
Start Time
Stop Time
T
T
T
Interval
Records for the specified recording
length
T
T
START
key
When the recording length exceeds
the specified interval
The next trigger is not applied until
the data for the specified recording
length has been acquired.
When the recording length exceeds
the stop time
Recording time depends on the operating
function.
When the interval is set to zero
If the
[Repeat]
trigger mode is
selected, or the REC or MEM functions
are used, recording is repeated from
Start to Stop times.
The timer trigger time and the time at which the trigger is actually fired may differ by up to 3 sample
times.
9
Trigger Settings
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......