Fine Adjustment of Input Values (Vernier Function)
8.7 Fine Adjustment of Input Values (Vernier
Function)
Fine adjustment of input voltage can be performed arbitrarily on the Waveform screen. When recording
physical values such as noise, temperature and acceleration using sensors, amplitude can be adjusted to
facilitate calibration.
Normal display
Vernier Function Enabled
1.2 V input is displayed as 1.0 V
1.0 V
1.2 V
Procedure
Press the
DISP
key
to open the Waveform screen, and then press the
CH.SET
key
to open the channel settings
window (
[Analog]
sheet).
1
Move the cursor to the [ ] vernier setting of the channel you
want to adjust.
2
Make the adjustment with the
F
keys while watching the
waveform.
Vernier↑
Magnify waveform
Vernier↓
Compress waveform
Vernier Reset
Return waveform to original position
•
The adjustment range is 50 to 200% of the original waveform. The magnification/compression ratio is
not displayed.
•
Vernier adjustments cannot be verified on printed waveforms or lists.
•
The waveform data (print data, saved file data) is adjusted by the Vernier function.
8
Utility Functions
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......