18.1 Trouble Shooting
366
18.1 Trouble Shooting
If damage is suspected, check the "Troubleshooting" section before
contacting your dealer or Hioki representative.
If Power and Operating Mouse Malfunction
Symptom
Check Item, or Cause
Remedy and Reference
The display does not
appear when you turn the
power on.
Is the power cord disconnected?
Are connections made correctly?
Verify that the power cord is connected
properly.
Cannot operate the instru-
ment with a mouse.
• Is the USB mouse connected?
• Are you using a hub or special
mouse?
• Is there a source of interference
nearby?
• Check the USB connector.
• If you are using a hub, connect the
mouse directly to the instrument.
Use an ordinary mouse.
• Move the mouse away from the source
of interference.
If the Display or Operations Malfunction
Symptom
Check Item, or Cause
Remedy and Reference
Data is not displayed on
an external LCD monitor.
• Is it connected properly?
• Is it a digital LCD monitor?
• Are you using a VGA-DVI adapter
cable?
• Check the connections and power of the
instrument and LCD monitor.
• An analog monitor (VGA output) cannot
be used. Use an LCD monitor with a DVI-
D (digital output) connector.
The monitor flickers.
• Is there a source of interference
nearby?
• Are you using a special LCD moni-
tor? (Although extremely uncom-
mon, flickering may be caused by
incompatibility with the LCD monitor.)
• Move the cable away from the source of
interference.
• Use an ordinary LCD monitor.
A waveform does not
appear when measure-
ment is started
• Is the “Pre-Trigger wait” message
displayed?
• Is the "Trigger wait" message dis-
played?
When pre-triggering is enabled, triggering is
ignored until the pre-trigger portion of the
waveform has been acquired. Recording
starts when a trigger occurs.
No changes occur in the
displayed waveform.
• Is the clamp sensor or connection
cable connected correctly?
• Is the vertical axis (voltage axis)
range set properly?
• Is the low-pass filter enabled?
Verify that the clamp sensor or connection
cable is connected correctly.
Verify the input channel settings.
While measuring with the
memory function, the dis-
played frequency is much
lower than the actual fre-
quency.
Aliasing may be occurring.
Change the timebase to use a faster sam-
pling rate.
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...
Страница 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Страница 302: ...12 8 FFT Analysis Modes 290 ...
Страница 316: ...13 6 Editor Command Details 304 ...
Страница 416: ...Appendix 4 FFT Definitions A26 ...
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......