13.2 Setting the Waveform Area
294
To evaluate the waveforms, a evaluation area is required. Two methods are available: one is to load the
already created evaluation area and settings, and the other is to create a new evaluation area.
Requirements for loading waveform evaluation area BMP files
Observe the following requirements when creating waveform evaluation areas on a computer and when load-
ing waveform evaluation areas created on other equipment:
Files that do not satisfy these requirements cannot be loaded properly.
13.2 Setting the Waveform Area
Loading the already created evaluation area
Setting method
Procedure Screen: FILE
1
Click
[FILE]
to call the File screen.
2
Select the media for loading.
3
Use the bar cursor to select the desired file.
4
Select
[Load]
.
5
Select
[Exec]
.
6
When no more changes need be made to the
loaded settings, press the
DISP
key to make the
Waveform screen appear, and then press the
START
key to initiate measurement.
To change the settings, first change the contents,
then press the
DISP
key to make the Waveform
screen appear, and then press the
START
key to
initiate measurement.
item
Condition
Loading operation
Color
White and black
White pixels are treated as background, and black pixels are
treated as the evaluation area.
Size
501 horizontal pixels
×
626 vertical pixels
The file will be used as the evaluation area for the Memory
Function 1 screen.
501 horizontal pixels
×
501 vertical pixels
When using the FFT function, the file will be used as the FFT
evaluation area. When using other functionality, the file will be
used as the memory function XY evaluation area.
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...
Страница 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Страница 302: ...12 8 FFT Analysis Modes 290 ...
Страница 316: ...13 6 Editor Command Details 304 ...
Страница 416: ...Appendix 4 FFT Definitions A26 ...
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......