11.2 Display Settings
242
Difference Between Dead Times During Normal and Memory Division
Recording
When Auto Save is set for continuous triggering
[Repeat]
Dead Times
Times during which sampling is inhibited due to internal processing or saving
Recording
Length
Anomalous phenomena occurring during dead times are not detected.
When the Trace Waveform Display is disabled (Off) during Memory Division
recording
1
2
3
4
5
6
Recording
Length
Dead Times
The waveform data of each recording length is recorded in one block.
When recording with Memory Division, dead time is shorter than with normal recording.
• The dead time (time where no sampling occurs between blocks) of memory division is as follows.
5
μ
s/div to 20
μ
s/div: 1 to 8 samples
Time axis slower than 50
μ
s/div: 1 samples
Note: Dead time lengthens during numerical calculation or waveform calculation, or when the
time axis is 5 to 20
μ
s/div and tracking waveform display is
[On]
.
• When measuring with an 8970 FREQ Unit, dead time is approximately 230 ms. When measuring
in integral value (
[Count]
) mode, there are cases when the last data in the previous block
remains in the first part of the block.
• When tracking waveform display is
[Off]
, the roll mode function cannot be used, even when roll
mode is enabled (not Off).
• When triggering occurs very often, clicking
[STOP]
may not stop measurement until enough data
has been acquired to fill the blocks specified for use.
Auto Save for Memory Division Recording
Measuring conditions
Auto Save
With Numerical calcula-
tion ON
Auto Save is performed every time one block is measured. When Tracking wave
display is ON, Waveform is also displayed.
When time axis is 5 to 20
μ
s/div with Tracking wave
display ON.
Auto Save and Waveform Display are performed every time one block is measured.
When time axis is 5 to 20
μ
s/div with Tracking wave
display OFF.
Auto Save is performed after all blocks are measured.
Other than above
Auto Save is performed simultaneously with measuring. When Tracking wave dis-
play is ON, Waveform is also displayed.
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...
Страница 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Страница 302: ...12 8 FFT Analysis Modes 290 ...
Страница 316: ...13 6 Editor Command Details 304 ...
Страница 416: ...Appendix 4 FFT Definitions A26 ...
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......