8.12 Using trigger settings to search measurement data
205
7
Cha
pter 8 T
rigger Set
tings
8
Trigger settings can be used to search measurement data.
Locations that match the set trigger criteria in the measurement data are searched for and displayed
sequentially.
Using the same method to set triggers, it is possible to search for criteria different to those at the
time of measurement.
Example 1: The measurement trigger criteria is trigger level 0 V and the data search condition is
level 100 V.
Example 2: After measurement with a free run without a trigger, locations that exceed 100 V are
searched for.
8.12 Using trigger settings to search
measurement data
Procedure
To open the screen: Right-click and select
[DISP]
→
Waveform screen
→
Right-click and select
[TRIG.SET]
→
Trigger settings window
(
[Analog Trig]
, or
[Logic Trig]
sheet)
1
After measurement is finished, set the search criteria.
The setting method is the same as described in "8.3 Trigger-
ing by Analog Signals" (p.188) and "8.4 Triggering by Logic
Signals (Logic Trigger)" (p.193).
2
Specify the search range.
Move the flashing cursor to the
[
Area
]
item.
Select
(When using memory division)
Select
3
Execute search.
Move the flashing cursor to the
[
Trig search
]
item.
Select
Whole Wave
Search all measured data. (default setting)
A-B Wave
Search in the range between A/B cursor.
Whole Display
Blocks
Applies search to the whole display blocks.
A/B Display
Blocks
When setting the A/B cursors on the display blocks, ap-
plies search to the range between the A and B cursors on
the display blocks.
Whole Blocks
Applies search to the data in the whole blocks.
A/B Whole
Blocks
When setting the A/B cursors for the whole blocks, applies
search to the range between the A and B cursors for the
whole blocks.
Execute
search
Start the search from the beginning of the measurement
data.
Search next
Searches for the next search condition at the most recent
search position.
Move A cursor
Move A cursor to the searched position.
Move B cursor
Move B cursor to the searched position.
Clear
Clear the searched position.
3
2
• When selecting A/B range and only
A cursor is shown on the screen,
search will be performed on the
data after A cursor.
• The search range for
[Whole
Blocks]
and
[A/B Whole Blocks]
will be defined according to the
start block and the number of
blocks to be measured.
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...
Страница 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Страница 302: ...12 8 FFT Analysis Modes 290 ...
Страница 316: ...13 6 Editor Command Details 304 ...
Страница 416: ...Appendix 4 FFT Definitions A26 ...
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......