7.4 Converting Input Values (Scaling Function)
148
The 9018-50 Clamp-On Probe provides 0.2 V output when measuring 10 A. So
Scaling should be set to display 10 A with 0.2 V input (and 0 A with 0 V input).
Selecting a Clamp Type ___________________________________________
1. Move the flashing cursor to the
[Clamp]
item, and select
[Select]
.
The flashing cursor moves to the
[Model]
item.
2. Select
[9000
]
.
The flashing cursor moves to the
[Clamp]
.
3. Select
[9018-50]
from the clamp list and select
[Confirm]
.
Units, scaling method, and ratio are set automatically.
4. Select the same range of the clamp when using the range selection type.
Select
[10A]
here.
To reset Scaling settings:
Move the flashing cursor to the
[Setting]
, and select
[Reset]
.
To copy the scaling setting to another channel
The Channel screen -
[Scaling]
sheet can be used to copy a setting.
See:
"7.8 Copying settings to other channels (calculation No.) (Copy function)" (p.158)
Using the Scaling and Variable functions (p.153) in combination:
The full span of output from a sensor can be displayed. (p.155)
At factory shipping, automatic correction of the variable function (p.307) is set to
[On]
.
At this time, the Variable setting is altered so that it is linked to (dependent upon)
the vertical axis (voltage axis) range and scaling settings. If you want the Vari-
able function setting to take priority, use either of the following procedures:
• Set Scaling first, and then set the Variable function
• Set a Variable value before Scaling, and then set Scaling.
When automatic correction of the Variable function (Variable Auto Adjustment) is
disabled (
[Off]
), the Scaling and Variable settings are unlinked (independent of
one another).
7.4.1 Scaling Setting Examples
Using a Clamp-On Probe
Example 1
Measure with the 10 A range of the Model 9018-10 Clamp-On Probe and
display the measured data in units of [A] (Amperes)
Setting Items
Setting Choice
Disp
NUM or SCI
Clamp
9018-50
Unit
*
A
Method
*
Ratio
Ratio
*
50.000
*:
Set automatically when clamp is selected.
Содержание MR8740
Страница 1: ...MR8740 MR8741 Instruction Manual MEMORY HiCORDER EN Oct 2018 Revised edition 7 MR8740A982 07 18 10H ...
Страница 2: ......
Страница 12: ...Contents x ...
Страница 26: ...Operating Precautions 14 ...
Страница 128: ...5 5 Miscellaneous Printing Functions 116 ...
Страница 196: ...7 12 Setting Output Waveform Parameters 184 ...
Страница 234: ...9 5 Numerical Calculation Type and Description 222 ...
Страница 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Страница 302: ...12 8 FFT Analysis Modes 290 ...
Страница 316: ...13 6 Editor Command Details 304 ...
Страница 416: ...Appendix 4 FFT Definitions A26 ...
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......