Automatically Specifying the Trigger Level (Auto-trigger Level)
5.12
Automatically Specifying the Trigger Level (Auto-
trigger Level)
When you enable this setting, the instrument performs a preliminary measurement before an actual
measurement.
In the preliminary measurement, the instrument acquires the specified number of data points and
calculates the average value of them. Then, the instrument specifies the average value as the
reference of the out trigger and performs the actual measurement using this trigger.
This function is useful when you use the out trigger setting with the trigger and alert functions.
1
2
> [Trigger] > [Common]
1
Click the
[Trigger]
to set it to
[On]
or
[Off]
.
Off
Disables the auto-trigger level function.
On
Automatically specifies the reference level of the out trigger.
2
Click the
[Samples]
box, and then choose an option for the number of sampling points from
the list.
100
,
200
,
300
,
400
,
500
3
Configure the trigger setting for each channel.
On the trigger source setting screen, set the trigger type of a channel to be used to the out trigger,
and set
[Auto-trigger level]
to
[On]
5
Configuring the
Trigger Settings
Содержание MR8740-50
Страница 2: ......
Страница 6: ...Contents iv...
Страница 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Страница 86: ...80 Configuring Generator Module Specific Settings...
Страница 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Страница 164: ...158 Numerical Calculation Types and Descriptions...
Страница 182: ...176 Displaying Waveforms...
Страница 186: ...180...
Страница 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Страница 224: ...218 External Sampling EXT SMPL...
Страница 232: ...226 Information for Reference Purposes...
Страница 236: ...230 Index...
Страница 237: ......
Страница 238: ......