
78
Configuring Module-Specific Settings
Configuring Model U8974 High Voltage Unit settings
> [Channel] > [U8974]
to
3
5
1
2
1
In the
[Mode]
area, tap the
[DC]
or
[RMS]
to choose a measurement mode.
DC
For voltage measurement
RMS
For RMS measurement
2
(When you have changed the measurement mode) Tap
[Zero adjust]
.
The instrument performs zero-adjustment. Execute zero-adjustment without any input.
3
Tap the area that includes
[Response]
.
The setting dialog box appears.
4
Tap the
[Response]
box, and then from the list, choose a response time for RMS
measurement.
Fast
Sets the response time to 150 ms.
Normal
Sets the response time to 500 ms.
Slow
Sets the response time to 2.5 seconds.
Setting the response time to
[Slow]
can stabilize measured values if frequency is relatively low or voltage
fluctuates severely.
5
Tap
[Close]
.
The setting dialog box closes.
HIOKI MR6000A966-03
Содержание Memory HiCorder MR6000
Страница 2: ...HIOKI MR6000A966 03...
Страница 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Страница 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Страница 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Страница 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Страница 391: ...HIOKI MR6000A966 03...
Страница 392: ...HIOKI MR6000A966 03...