66
Configuring Module-Specific Settings
Configuring Model 8967 Temp Unit settings
> [Channel] > [8967]
to
2
5
1
1
Tap the
[Mode]
box, and then choose a thermocouple type from the list.
Choose an option depending on the type of a thermocouple to be used.
Mode
Measurable range
Mode
Measurable range
K
−200°C to 1350°C
R
0°C to 1700°C
J
−200°C to 1100°C
S
0°C to 1700°C
E
−200°C to 800°C
B
400°C to 1800°C
T
−200°C to 400°C
W
0°C to 2000°C
N
−200°C to 1300°C
2
Tap the area that includes
[RJC]
.
The setting dialog box appears.
3
Tap the
[RJC]
box, and then choose between
[Int.]
and
[Ext.]
from the list for reference
junction compensation.
Int.
Execute the reference junction compensation inside the module.
(Measurement accuracy: The sum of the accuracy of the temperature measurement
and that of the reference junction compensation)
Ext.
Does not execute the reference junction compensation inside the module.
(Measurement accuracy: The accuracy of the temperature measurement only)
When connecting a thermocouple directly to the module, choose
[Int.]
.
When connecting a thermocouple via a reference junction device (e.g., zero-point bath), choose
[Ext.]
.
HIOKI MR6000A966-03
Содержание Memory HiCorder MR6000
Страница 2: ...HIOKI MR6000A966 03...
Страница 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Страница 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Страница 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Страница 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Страница 391: ...HIOKI MR6000A966 03...
Страница 392: ...HIOKI MR6000A966 03...