361
Information for Reference Purposes
Maximum
recordable time when the real-time saving is enabled (values
for reference purposes)
The maximum recordable time is expressed in the following equation.
(Maximum recordable time) = [(Recording capacity) × (Sampling time)] ÷
[(Number of channels used) × 2]
(Number of channels used) = [(Number of analog channels other than Model MR8990)
+ (Number of logic modules) + (Number of Model MR8990 channels)] × 2
+ (Number of real-time calculation channels) × 2
The maximum recordable times for saving data to each storage device are shown in the following
table (assume that each storage device is empty). Since no capacity of the header of a waveform
file is included, use about 90% of the recordable time provided in the table as a reference.
Some conditions allow a long-term recordable time (1 year or more, shaded areas in the tables) to
be set; however, the operation cannot be guaranteed because the warranty period or product life
may disturb it.
For saving data acquired without using the envelope on Model U8332 SSD Unit
d: days, h: hours, min: minutes, s: seconds
Sampling
rate
Number of channels used
2
4
8
16
32
20 MS/s
53 min 20 s
–
–
–
–
10 MS/s
1 h 46 min 40 s
53 min 20 s
–
–
–
5 MS/s
3 h 33 min 20 s
1 h 46 min 40 s
53 min 20 s
–
–
2 MS/s
8 h 53 min 20 s
4 h 26 min 40 s
2 h 13 min 20 s
1 h 6 min 40 s
–
1 MS/s
17 h 46 min 40 s
8 h 53 min 20 s
4 h 26 min 40 s
2 h 13 min 20 s
1 h 6 min 40 s
500 kS/s
1 d 11 h 33 min 20 s
17 h 46 min 40 s
8 h 53 min 20 s
4 h 26 min 40 s
2 h 13 min 20 s
200 kS/s
3 d 16 h 53 min 20 s
1 d 20 h 26 min 40 s
22 h 13 min 20 s
11 h 6 min 40 s
5 h 33 min 20 s
100 kS/s
7 d 9 h 46 min 40 s
3 d 16 h 53 min 20 s
1 d 20 h 26 min 40 s
22 h 13 min 20 s
11 h 6 min 40 s
50 kS/s
14 d 19 h 33 min 20 s
7 d 9 h 46 min 40 s
3 d 16 h 53 min 20 s
1 d 20 h 26 min 40 s
22 h 13 min 20 s
20 kS/s
37 d 0 h 53 min 20 s
18 d 12 h 26 min 40 s
9 d 6 h 13 min 20 s
4 d 15 h 6 min 40 s
2 d 7 h 33 min 20 s
10 kS/s
74 d 1 h 46 min 40 s
37 d 0 h 53 min 20 s
18 d 12 h 26 min 40 s
9 d 6 h 13 min 20 s
4 d 15 h 6 min 40 s
5 kS/s
148 d 3 h 33 min 20 s
74 d 1 h 46 min 40 s
37 d 0 h 53 min 20 s
18 d 12 h 26 min 40 s
9 d 6 h 13 min 20 s
2 kS/s
370 d 8 h 53 min 20 s
185 d 4 h 26 min 40 s
92 d 14 h 13 min 20 s
46 d 7 h 6 min 40 s
23 d 3 h 33 min 20 s
1 kS/s
740 d 17 h 46 min 40 s
370 d 8 h 53 min 20 s
185 d 4 h 26 min 40 s
92 d 14 h 13 min 20 s
46 d 7 h 6 min 40 s
500 S/s 1481 d 11 h 33 min 20 s
740 d 17 h 46 min 40 s
370 d 8 h 53 min 20 s
185 d 4 h 26 min 40 s
92 d 14 h 13 min 20 s
200 S/s 3703 d 16 h 53 min 20 s 1851 d 20 h 26 min 40 s
925 d 22 h 13 min 20 s
462 d 23 h 6 min 40 s
231 d 11 h 33 min 20 s
100 S/s
7407 d 9 h 46 min 40 s 3703 d 16 h 53 min 20 s 1851 d 20 h 26 min 40 s
925 d 22 h 13 min 20 s
462 d 23 h 6 min 40 s
50 S/s
10000 d
7407 d 9 h 46 min 40 s 3703 d 16 h 53 min 20 s 1851 d 20 h 26 min 40 s
925
d
22
h
13 min 20
s
20 S/s
10000 d
10000 d
9259 d 6 h 13 min 20 s
4629
d
15
h
6 min 40
s 2314
d
19
h
33
min
20
s
10 S/s
10000 d
10000 d
10000 d
9259
d
6
h
13 min 20
s
4629
d
15
h
6 min
40
s
5 S/s
10000 d
10000 d
10000 d
10000 d
9259
d
6
h
13 min 20
s
2 S/s
10000 d
10000 d
10000 d
10000 d
10000 d
1 S/s
10000 d
10000 d
10000 d
10000 d
10000 d
Appendix
HIOKI MR6000A966-03
Содержание Memory HiCorder MR6000
Страница 2: ...HIOKI MR6000A966 03...
Страница 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Страница 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Страница 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Страница 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Страница 391: ...HIOKI MR6000A966 03...
Страница 392: ...HIOKI MR6000A966 03...