322
Managing Data in the Instrument With the FTP Server Function
14.2
Managing Data in the Instrument With the FTP
Server Function
Using the FTP client software installed in your computer allows you to transfer files from storage
devices in the instrument to the computer, and manage the files with the computer.
• The instrument is equipped with an FTP server.
• You can operate the instrument using Internet Explorer
®
or various free software.
Precautions when the FTP is used
• The FTP server of the instrument allows only one connection. Multiple computers cannot
simultaneously access the server.
• If no command is sent from the computer for more than one minute after the FTP connection has
been established, the FTP session may time out. If a timeout occurs, reconnect the computer via
the FTP.
• The instrument interrupts an FTP session during real-time saving (during measurement).
•
Before inserting or removing an SD memory card or a USB flash drive, terminate the FTP
connection first.
•
Do not handle any files during an FTP session.
•
With Internet Explorer
®
, updated dates of files may not match those saved in the instrument.
•
With Internet Explorer
®
, temporary Internet files created during the previous access are retained;
thus, the previous data may be acquired instead of the latest data.
•
If you cancel the transfer of selected files or folders before its completion, some FTP client/
browser programs may delete all of them from the source regardless of whether they have been
transferred. Avoid moving files and folders. Download files and folders using the copy function,
and then delete them.
• Be careful that the user name and password themselves will get transmitted via the network when
the FTP server is used.
•
When you upload a file, the time stamp on the file will change to the uploading time.
• Each storage device appears as a directory on the FTP server.
/STORAGE
Waveform data written in the instrument (Writing data is prohibited.)
/D
through
/Z
Storage devices serving as drives
HIOKI MR6000A966-03
Содержание Memory HiCorder MR6000
Страница 2: ...HIOKI MR6000A966 03...
Страница 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Страница 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Страница 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Страница 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Страница 391: ...HIOKI MR6000A966 03...
Страница 392: ...HIOKI MR6000A966 03...