296
Configuring the CAN/LIN Definition Data Settings
Data allocation (
[Byte order]
)
Captured data using a big-endian ordering differs from those using a little-endian ordering although
the same data starting address (
[Start bit]
) and the same bit length (
[Bit length]
) are uses.
In the following figure, the shaded cells indicate the pieces of data to be captured when a bit length
of 10 and a data starting address (LSB position) of bit 16 are configured.
Big-endian ordering
The MSB is configured at address 9.
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
7
6
5
4
3
2
1
0
Byte 1
15
14
13
12
11
10
9
8
Byte 2
23
22
21
20
19
18
17
16
Byte 3
31
30
29
28
27
26
25
24
Little-endian ordering
The MSB is configured at address 25.
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
7
6
5
4
3
2
1
0
Byte 1
15
14
13
12
11
10
9
8
Byte 2
23
22
21
20
19
18
17
16
Byte 3
31
30
29
28
27
26
25
24
In the following figure, assuming that byte 0 and byte 1 have 16-bit data of FF and 00, respectively.
The following 16-bit data sets are captured:
Big-endian ordering (MSB first): FF00 (Data starting address: 8)
Little-endian ordering (LSB first): 00FF (Data starting address: 0)
First of the data frame
Toward the last of the data frame
7 6 5 4 3 2 1 0 7 6 5 4 3 2 1 0 7 6 5 4 3 2 1 0 7 6 5 4 . . .
Byte 0
(1st byte)
Byte 1
(2nd byte)
Byte 2
(4th byte)
Byte 3
(2nd byte)
HIOKI MR6000A966-03
Содержание Memory HiCorder MR6000
Страница 2: ...HIOKI MR6000A966 03...
Страница 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Страница 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Страница 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Страница 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Страница 391: ...HIOKI MR6000A966 03...
Страница 392: ...HIOKI MR6000A966 03...