Configuring Dual Sampling Settings
4
Set the sampling speed and the number of points for instantaneous waveforms.
The instrument measures instantaneous waveforms using the normal method (acquiring one data point per
sample). Choose a sampling rate faster than that chosen for trend waveforms. See the following page for the
sampling speeds that can be set.
See “Linear time-invariant system” (p. 372).
5
Configure the memory division settings.
Instantaneous waveforms can be acquired into multiple blocks.
By setting the dual sampling to
[On]
, the memory division is also set to
[On]
automatically, allowing you to set
the number of blocks, start block number, and the number of blocks to be used.
See “10 Memory Division Function” (p. 255).
6
Configure the input channel settings.
> [Channel]
Configure the analog and logic channels settings.
See “1.3 Configuring the Input Channel Settings” (p. 11).