233
Setting FFT Calculation Conditions
Configuring the calculation result peak value settings
The instrument can display global or local maximum values of input signals and calculation results
on the waveform screen.
> [Calculation] > [FFT calculation]
1
2
3
1
Tap the
[FFT]
button to set it to
[On]
.
2
Tap the open mark under
[Detail]
.
The FFT calculation advanced setting window appears.
3
Tap the
[Peak]
box, and then from the list, choose a peak value type to be displayed.
Off
Does not display any peak values.
2
3
2
4
5
6
1
1
Maximal value
Maximum value
Maximal
Displays the ten greatest values among
maximal values. When a value of a data
point is greater than values of both the
adjacent points, the instrument determines
that the point is a local maximum.
Maximum
Displays the ten greatest values.
•
When the instrument does not find any peak values, it displays no values.
•
The instrument can display peak values on the waveform screen, but no peak values can be
stored in text-form files.
•
Some split screens does not always display ten peak values. Such a screen displays as many
values as possible in decreasing order.
FFT
Calculation Function
HIOKI MR6000A966-03
Содержание Memory HiCorder MR6000
Страница 2: ...HIOKI MR6000A966 03...
Страница 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Страница 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Страница 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Страница 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Страница 391: ...HIOKI MR6000A966 03...
Страница 392: ...HIOKI MR6000A966 03...