142
Externally Triggering the Instrument (External Trigger)
5.9 Externally Triggering the Instrument
(External Trigger)
External signals applied to the external control terminals can serve as trigger sources.
External signals can also be used to operate multiple instruments in synchronization with each
other.
See “External trigger terminal (EXT.TRIG)” (p. 353).
5.10 Manually Triggering the Instrument
(
Forcible Trigger)
Tapping
[Trigger]
on the right side of the waveform screen allows you to manually trigger the
instrument while the instrument is waiting for a trigger.
The forcible trigger triggers the instrument regardless of other trigger source settings.
To stop the recording, press the
STOP
key.
Press the key once: Stops the measurement once the instrument has acquired the specified
recording length of data.
Press the key twice: Stops the recording immediately.
HIOKI MR6000A966-03