137
Triggering the Instrument Using Analog Signals (Analog Trigger)
5.
[Glitch]
trigger
The glitch-trigger condition is satisfied when a pulse width of an input signal that has crossed the
specified level is shorter than the specified width. These triggers are disabled when the sampling
rate is set at 200 MS/s.
The external sampling is cannot be set when the envelope is used. This trigger cannot be set when
Model MR8990 is used.
Width
Level
Input waveform
Slope:
[ ]
Setting
Description
Level
–f.s.
to
+f.s.
Default:
0
Allows you to specify a level for detecting glitches.
Slope
,
Allows you to choose which of the following points to use
to detect glitches: two consecutive points at which a signal
crosses the specified level while rising; or those while falling.
Event
With
OR
1
to
4,000
Allows you to enter the number of events.
The instrument counts the number of times the glitch-trigger
condition is satisfied. An analog trigger is generated only
after the count reaches the specified number of events.
With
AND
Not available
Width
2
times to
4000
times of
the sampling period
Allows you to enter a pulse width (time), which is used to
determine a glitch.
The glitch-trigger condition is satisfied when a pulse width is
shorter than the specified width. (The available setting range
varies depending on the sampling periods. Lower limit: 2
times sampling period or more; Upper limit: 4000 times
sampling period or less)
Configuring the
Trigger Settings
HIOKI MR6000A966-03
Содержание Memory HiCorder MR6000
Страница 2: ...HIOKI MR6000A966 03...
Страница 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Страница 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Страница 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Страница 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Страница 391: ...HIOKI MR6000A966 03...
Страница 392: ...HIOKI MR6000A966 03...