Triggering the Instrument Using Analog Signals (Analog Trigger)
2.
[In]
trigger,
[Out]
trigger
When an input signal falls within (
[In]
) or exits of a range (
[Out]
), which is determined by specifying
upper and lower values, an analog trigger is generated. These triggers are disabled when the
sampling rate is set at 200 MS/s.
[In]
Upper limit value
Lower limit value
Upper limit value
Lower limit value
[Out]
Setting
Description
Event
With
OR
1
to
4,000
Allows you to enter the number of events.
The instrument counts the number of times the window-trigger
condition is satisfied. Only after the count reaches the specified
number of events, an analog trigger is generated.
With
AND
Not available
Filter
Off
,
10
to
10,000
Allows you to enter a filter width in the number of samples.
Only after the window-trigger condition is continuously satisfied
during the specified period, the analog trigger is generated.
Specifying this option prevents the instrument from being
unintentionally triggered due to noise.
Off
,
1 ms
,
10 ms
Allows you to enter a filter width in milliseconds when the
envelope is used.
Upper
–f.s.
to
+f.s.
Allows you to enter an upper limit value.
Lower
–f.s.
to
+f.s.
Allows you to enter a lower limit value.
Behavior with the in trigger and out trigger settings varies depending on the trigger satisfaction
conditions (AND and OR operations).
With
OR
In
The window-trigger condition is satisfied when an input signal
crosses the upper or lower limit value of the threshold (level)
and thereby falls within the range, which is specified by the
upper and lower limit values of the threshold (level).
Out
The window-trigger condition is satisfied when an input signal
crosses the upper or lower limit value of the threshold (level)
and thereby exits the range, which is specified by the upper
and lower limit values of the threshold (level).
With
AND
In
The window-trigger condition is satisfied when an input signal
is inside the range, which is specified by the upper and lower
limit values of the threshold (level).
Out
The window-trigger condition is satisfied when an input signal
is outside the range, which is specified by the upper and lower
limit values of the threshold (level).
Configuring the
Trigger Settings
HIOKI MR6000A966-03
Содержание Memory HiCorder MR6000
Страница 2: ...HIOKI MR6000A966 03...
Страница 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Страница 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Страница 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Страница 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Страница 391: ...HIOKI MR6000A966 03...
Страница 392: ...HIOKI MR6000A966 03...