131
Triggering the Instrument Using Analog Signals (Analog Trigger)
1.
[Level]
trigger
When an input signal crosses the specified level in the positive or negative direction, an analog
trigger is generated.
Level
Input waveform
Slope:
[ ]
[ ]
Setting
Description
Level
–f.s.
to
+f.s.
Default:
0
Allows you to enter a level of the level trigger.
Slope
With
OR
The level-trigger condition is satisfied when a waveform crosses
the threshold value (level) while rising.
The level-trigger condition is satisfied when a waveform crosses
the threshold value (level) while falling.
With
AND
HIGH
The level-trigger condition is satisfied when a waveform is higher
than the threshold value (level).
LOW
The level-trigger condition is satisfied when a waveform is lower
than the threshold value (level).
Event
With
OR
1
to
4,000
Allows you to enter the number of events.
The instrument counts the number of times the level-trigger
condition is satisfied. An analog trigger is generated when the
count reaches the specified number of events.
With
AND
Not available
Filter
Off
,
10
to
10,000
Allows you to enter a filter width in the number of samples. Only
after the level-trigger condition is continuously satisfied during
the specified period, an analog trigger is generated. Specifying
this option prevents the instrument from being unintentionally
triggered due to noise.
Off
,
1 ms
,
10 ms
Allows you to enter a filter width in milliseconds when the
envelope is used.
With the
[Filter]
setting
Specifying the filter width prevents analog triggers from being unintentionally generated due to
noise, allowing it to be generated only after the level-trigger condition is continuously satisfied
during the specified width or longer.
Filter
Level
Noise does not generate any analog trigger.
Configuring the
Trigger Settings
HIOKI MR6000A966-03
Содержание Memory HiCorder MR6000
Страница 2: ...HIOKI MR6000A966 03...
Страница 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Страница 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Страница 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Страница 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Страница 391: ...HIOKI MR6000A966 03...
Страница 392: ...HIOKI MR6000A966 03...