127
Setting the Trigger Satisfaction Conditions (AND/OR Operation) Among Trigger Sources
5.5 Setting the Trigger Satisfaction Conditions
(
AND/OR Operation) Among Trigger Sources
Set the trigger satisfaction conditions among the analog, logic, CAN, external, and interval triggers
by choosing between logical AND or OR operation.
The forcible trigger triggers the instrument regardless of the trigger satisfaction conditions setting. If
all trigger sources are set to off (i.e., with no trigger setting), recording starts immediately (free run).
> [Trigger] > [Common]
Tap
[AND]
or
[OR]
, whichever is displayed, to switch to the other.
OR
When any one of the specified trigger conditions changes from not being satisfied to
being satisfied (at a changing point), the instrument is triggered. Thus, even though
a trigger condition has been already satisfied at the start of waiting for a trigger, the
instrument is not triggered until a changing point is detected.
AND
Only when all of the specified trigger conditions are satisfied, the instrument is
triggered. Thus, if all the specified trigger conditions have been already satisfied at the
start of waiting for a trigger, the instrument is immediately triggered.
Configuring the
Trigger Settings
HIOKI MR6000A966-03
Содержание Memory HiCorder MR6000
Страница 2: ...HIOKI MR6000A966 03...
Страница 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Страница 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Страница 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Страница 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Страница 391: ...HIOKI MR6000A966 03...
Страница 392: ...HIOKI MR6000A966 03...