6
Setting Measurement Conditions
3
Tap the
[External sampling]
button to set it to
[On]
or
[Off]
.
The external sampling is cannot be set when the envelope is used.
Off
Disables the external sampling function.
On
Choose this option to sample data at a sampling rate defined by a signal inputted into
the external sampling terminal (EXT.SMPL).
Samples data at rising edges of the input signal.
Samples data at falling edges of the input signal.
&
Samples data at both rising and falling edges of the input signal.
4
Tap the
[Points]
box, and then from the list, choose an option for the number of points to be
measured.
2.5 k
,
5 k
,
10 k
,
20 k
,
50 k
,
100 k
,
200 k
,
500 k
,
1 M
,
2 M
,
5 M
,
10 M
,
20 M
,
50 M
,
100 M
,
200 M
,
500 M
,
1 G
Tap
[Any]
to enable it, and then tap
[Points]
, which allows you to enter the number of points in 100
increments.
When the real-time saving is set to
[On]
, you cannot specify the
recording length in the
[Points]
box. Choose
[Save]
, and enter the recording length in the
[Recording time]
The maximum recording length you can set varies depending on the number of channels to be measured.
5
Tap the
[Mode]
box, and then choose a recording mode from the list.
Single
Measures waveforms only once. Pressing the
START
key starts recording waveforms,
and then stops when the specified recording length of the waveforms have been
acquired.
Repeat
Measures waveforms repeatedly. Pressing the
STOP
key stops the measurement in
progress.
When you choose
[Repeat]
and set
[Count]
to
[On]
, measurements repeat the number of times entered in
the
[Count]
box.
When
[Realtime save]
is set to
[On]
, you can choose
[Single]
mode only.
6
Configure the overlay setting.
See “3.1 Overlaying New Waveforms With Previously Acquired Waveforms” (p. 52).
7
Tap the
[Realtime save]
button to set it to
[On]
.
The instrument can record data in an optional storage device while measuring waveforms.
See “Real-time saving” (p. 106).
8
Configuring the memory division settings
The instrument can divide the internal memory into several blocks and store waveforms into them.
See “10 Memory Division Function” (p. 255).
HIOKI MR6000A966-03
Содержание Memory HiCorder MR6000
Страница 2: ...HIOKI MR6000A966 03...
Страница 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Страница 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Страница 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Страница 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Страница 391: ...HIOKI MR6000A966 03...
Страница 392: ...HIOKI MR6000A966 03...