5
Setting Measurement Conditions
1.2 Setting Measurement Conditions
You have to set conditions required for measurement, such as the sampling rate (
[Sampling]
) and
recording length (
[Shot]
).
> [Status] > [Condition]
7
1
3
6
8
5
2
4
1
Enter a comment in the
[Title comment]
box.
Number of characters that can be entered: up to 40
2
Tap the
[Sampling]
box, and then choose a sampling rate from the list.
See “Sampling rate setting guideline” (p. 7).
200 MS/s
,
100 MS/s
,
50 MS/s
,
20 MS/s
,
10 MS/s
,
5 MS/s
,
2 MS/s
,
1 MS/s
,
500 kS/s
,
200 kS/s
,
100 kS/s
,
50 kS/s
,
20 kS/s
,
10 kS/s
,
5 kS/s
,
2 kS/s
,
1 kS/s
,
500 S/s
,
200 S/s
,
100 S/s
,
50 S/s
,
20 S/s
,
10 S/s
,
5 S/s
,
2 S/s
,
1 S/s
The instrument can measure signals at a sampling rate of 200 MS/s even with several modules other than
Model U8976 installed together. However, the data refresh rate does not exceed the maximum sampling rate
of each module.
When the real-time waveform calculation (Model MR6000-01 only) is set to
[On]
, you cannot choose a
sampling rate of 200 MS/s.
When the real-time saving is set to
[On]
, due to the combination of the number of channels to be used and
save destinations, the maximum sampling rate that can be set varies as follows:
Number of channels to be used
Maximum sampling rate that can be chosen
SSD
HDD
USB flash drive
SD Card
FTP transmission
1 channel to 2 channels
20 MS/s
10 MS/s
5 MS/s
3 channel to 4 channels
10 MS/s
5 MS/s
2 MS/s
5 channel to 8 channels
5 MS/s
2 MS/s
1 MS/s
9 channel to 16 channels
2 MS/s
1 MS/s
500 kS/s
17 channel to 32 channels
1 MS/s
500 kS/s
200 kS/s
Only if the following Hioki-designated options are used, the real-time saving operation with the instrument is
guaranteed:
• Model U8332 SSD Unit
Model U8333 HD Unit
• Model Z4006 USB Drive
Model Z4001 and Model Z4003 SD Memory Card
Measurement Method
HIOKI MR6000A966-03
Содержание Memory HiCorder MR6000
Страница 2: ...HIOKI MR6000A966 03...
Страница 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Страница 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Страница 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Страница 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Страница 391: ...HIOKI MR6000A966 03...
Страница 392: ...HIOKI MR6000A966 03...