34
Configuring Input Channels
Measuring logic signals
This section describes how to configure settings on the individual settings screen when measuring
logic signals.
You can use the settings list screen to configure the settings. (See p. 52.)
External control terminals: Pulse input terminals P1 to P8
>
>
4
4
1
1
3
3
5
5
2
2
1
Select
[Pulse]
, and choose a channel from
[P1]
to
[P8]
,
Measurement will not be performed for channels whose checkboxes are not selected.
2
Select the waveform display color.
×
(OFF), 24 colors
Select
[×]
if you wish to measure the channel but not to display its waveform on the screen.
3
Set the input type to
[Logic]
.
4
Under
[Threshold]
, select the level to count.
1 V
Treats voltages that are greater than or equal to 1.0 V as high level and voltages that
are greater than or equal to 0 V but less than 0.5 V as low level.
4 V
Treats voltages that are greater than or equal to 4.0 V as high level and voltages that
are greater than or equal to 0 V but less than 1.5 V as low level.
5
Under
[Filter]
, select whether to use the chatter prevention filter.
Set to
[ON]
to prevent false counting due to chatter in mechanical contact (relay) output.
OFF
,
ON
Содержание LR8450
Страница 68: ...64 Observing Waveforms Alarm...
Страница 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Страница 131: ...128 Acquiring Data with a Computer PC...
Страница 175: ...172 Controlling the System...
Страница 233: ...230 Controlling the Instrument with Communication Commands...
Страница 271: ...268 Plug in Module Specifications...