Appendix 1 Error Messages and Remedial Actions
A
1
Appendix
Error messages consist of either "Error" or "Warning" displays. A screen message
appears whenever an error occurs. In either case, take the remedial action indicated.
To clear an error message, press the
ENTER
or
ESC
key.
Appears just once when an error occurs. Disappears within a few seconds.
Also disappears when any key is pressed.
Appendix
Appendix 1 Error Messages and Remedial
Actions
Error Messages
Message
Remedial Action
004
Battery low. Cannot access to the
waveform backup and Media.
Low battery charge: recharge or replace with
a fresh battery pack.
005
FPGA error.
Incompatible FPGA firmware version. Try up-
dating the firmware, and if the error persists,
submit the Memory HiLogger for repair.
013
Failure has been detected on power
source for USB memory.
Replace the USB flash drive. If the error per-
sists after replacement, the instrument needs
to be repaired.
014
Storage media has been removed
without going through a proper
[EJECT] process.
When removing storage media during auto
save operation, be sure to perform the [Eject]
procedure.
Warning Messages
Message
Remedial Action
101
Invalid key.
The pressed key is disabled due to measuring
or other operation.
Press the key again after the current operation
finishes.
102
Cannot start measurement.
The START/STOP key is disabled when the cur-
rent screen is displayed. Switch to the Wave-
form screen before starting measurement.
103
Cannot change while measuring.
Press the START/STOP key to stop measuring
before changing the setting.
Содержание LR8431
Страница 1: ...EN LR8431 20 MEMORY HiLOGGER Instruction Manual Aug 2015 Revised edition 3 LR8431B980 03 15 08H...
Страница 2: ......
Страница 20: ...Operating Precautions 12...
Страница 44: ...2 5 Zero Adjustment 36...
Страница 74: ...3 7 Viewing All Channel Settings in a List 66...
Страница 88: ...4 5 About Recording Operation 80...
Страница 120: ...6 6 Transferring Data to a Computer USB Drive Mode 112...
Страница 126: ...7 2 Numerical Value Calculation Expressions 118...
Страница 136: ...8 3 Making System Settings 128...
Страница 144: ...9 5 Synchronous Measurements with Multiple Instruments 136...
Страница 154: ...Chapter 10 Specifications 146...
Страница 198: ...Appendix 11 Pulse Input Circuit Diagram A40...
Страница 202: ...Index Index 4...
Страница 203: ......
Страница 204: ...EN LR8431 20 MEMORY HiLOGGER Instruction Manual Aug 2015 Revised edition 3 LR8431B980 03 15 08H...