5.9 Alarm Output
158
Alarms can be output at thermocouple burnout, communications
interruption, and low remaining battery life.
1
Select the
[Trig & Alm]
Screen.
2
Select a
[Alarm]
.
3
Select the criterion for which you wish to output an alarm and set it to on.
1
2
3
Set
[Alarm at burn out.]
to
[ON]
.
Alarm output will be generated when a burnout is detected, regardless of the
[AND]
and
[OR]
settings for other alarm criteria settings and alarm source set-
tings.
Set
[Alarm at disconnect.]
to
[ON]
.
A warning is output when the instrument is unable to communicate with a mea-
surement unit or wireless logger, regardless of the
[AND]
and
[OR]
settings for
other alarm criteria settings and alarm source settings.
Set
[Alarm at low battery]
to [ON].
A warning is output when a measurement unit or wireless logger indicates low
battery strength, regardless of the
[AND]
and
[OR]
settings for other alarm cri-
teria settings and alarm source settings.
Set
[Alarm at disconnect.]
to
[DELAY]
.
If the data containing 2000 or more samples is not received from the unit with
[Alarm at disconnect.]
set to
[DELAY]
, the alarm output will be generated.
Содержание LR8410-20
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents x...
Страница 32: ...Operating Precautions 20...
Страница 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Страница 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Страница 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Страница 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Страница 316: ...13 3 Disposing of the instrument 304...
Страница 360: ...Index Index 4...
Страница 361: ......
Страница 362: ......
Страница 363: ......
Страница 364: ......