7.8 Processing Recorded Data
109
3
7
Cha
p
ter 7
Using t
h
e LR500
0 Utili
ty Program
Dew-point temperature is calculated from the temperature and humidity measurement
data from the logger.
Calculation results are saved as a new item in the recording file.
1. Select the items and time span.
2. Click the
[Execute]
button.
(Calculation results are saved.)
Note: Click the
[Finish]
button to close the
[Dew Point]
dialog box.
Calculating Dew-Point Temperature
• There is no way to confirm that a specified data item is really a tem-
perature or humidity value. Dew-point calculation occurs regardless of
data type.
• Only the specified temperature and humidity data measured during
the specified recording time span is applied to calculations and saved.
• The valid range for calculation input measurement data is -100 to 100
degrees, and 0 to 100% humidity. Values outside of these ranges are
replaced with the maximum or minimum value within the valid range.
Item and range settings
Specify the temperature and humidity values,
and the time span for calculation.
1
2
Содержание LR5092
Страница 2: ......
Страница 14: ...Preparation for Use to Data Analysis 8...
Страница 36: ...3 6 Starting and Stopping Recording on Logger 30...
Страница 44: ...4 1 Collecting Recorded Data of Logger 38...
Страница 140: ...9 4 Disposing of the Logger 134...
Страница 144: ...Index Index iv...
Страница 145: ...HIOKI E E CORPORATION...
Страница 146: ......
Страница 147: ......
Страница 148: ......