7.2 Measurement Specifications
152
7.2.4 Temperature Measurement
Temperature and humidity range for guaranteed accuracy:
23±5°C less than 80% RH (no condensation)
Accuracy when using with the 9631 TEMPERATURE SEN-
SOR
When the 9631-05 TEMPERATURE SENSOR is used, the accuracy is
guaranteed within 0.0°C to 40.0°C.
Measuring range
Measurement accuracy
-10.0°C to -0.1°C
±1.5°C
0.0°C to 40.0°C
±1.0°C
40.1°C to 70.0°C
±1.5°C
Measurement Range, Accuracy
Temperature
characteristics
Measurement accuracy × 0.5 is added to
the measurement accuracy. (when the
ambient temperature is not 23±5°C)
Response time
Approx. 100 sec. Including the period of
time for the response of the 9631-01,
9631-05 TEMPERATURE SENSORs.
(Reference value: Period of time until
90% of the change in temperature is
reflected in the indication)
Influence of
radioactive RF
electromagnet-
ic field
±2°C at 3V/m
Содержание IR3455
Страница 2: ......
Страница 24: ...Operating Precautions 18...
Страница 110: ...4 4 Step Voltage Test 104...
Страница 160: ...7 3 9750 01 02 03 11 12 13 and 9751 01 02 03 ALLI GATOR CLIPs Specifications 154...
Страница 180: ...Appendix 4 Temperature Compensation Table 174...
Страница 181: ...1 2 3 4 5 6 7 8 9 10 11...
Страница 182: ......
Страница 183: ......
Страница 184: ......